MicroDicom DICOM Viewer
1. EXECUTIVE SUMMARY
- CVSS v4 5.7
- ATTENTION: Low attack complexity
- Vendor: MicroDicom
- Equipment: DICOM Viewer
- Vulnerability: Improper Certificate Validation
2. RISK EVALUATION
Successful exploitation of this vulnerability could allow an attacker to alter network traffic and perform a machine-in-the-middle (MITM) attack.
3. TECHNICAL DETAILS
3.1 AFFECTED PRODUCTS
The following MicroDicom DICOM Viewer are affected:
- MicroDicom DICOM Viewer: Version 2024.03
3.2 VULNERABILITY OVERVIEW
3.2.1 IMPROPER CERTIFICATE VALIDATION CWE-295
MicroDicom DICOM Viewer fails to adequately verify the update server's certificate, which could make it possible for attackers in a privileged network position to alter network traffic and carry out a machine-in-the-middle (MITM) attack. This allows the attackers to modify the server's response and deliver a malicious update to the user.
CVE-2025-1002 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N).
A CVSS v4 score has also been calculated for CVE-2025-1002. A base score of 5.7 has been calculated; the CVSS vector string is (AV:A/AC:L/AT:P/PR:N/UI:A/VC:N/VI:H/VA:N/SC:N/SI:N/SA:N).
3.3 BACKGROUND
- CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
- COUNTRIES/AREAS DEPLOYED: Worldwide
- COMPANY HEADQUARTERS LOCATION: Bulgaria
3.4 RESEARCHER
Sharon Brizinov of Claroty Team82 reported this vulnerability to CISA.
4. MITIGATIONS
MicroDicom recommends users upgrade to DICOM Viewer version 2025.1.
CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability, such as:
- Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.
- Locate control system networks and remote devices behind firewalls and isolating them from business networks.
- When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.
CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.
Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.
No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.
5. UPDATE HISTORY
- February 6, 2025: Initial Publication