ICS Medical Advisory

Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Application

Release Date
Alert Code
ICSMA-25-058-01

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 8.7
  • ATTENTION: Exploitable remotely/low attack complexity
  • Vendor: Dario Health
  • Equipment: USB-C Blood Glucose Monitoring System Starter Kit Android Application, Application Database and Internet-based Server Infrastructure
  • Vulnerabilities: Exposure of Private Personal Information to an Unauthorized Actor, Improper Output Neutralization For Logs, Storage of Sensitive Data In a Mechanism Without Access Control, Cleartext Transmission of Sensitive Information, Cross-site Scripting (XSS), Sensitive Cookie Without 'HttpOnly' Flag, Exposure of Sensitive Information Due To Incompatible Policies

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to expose information, inject code, manipulate data, or achieve cross-site scripting (XSS), resulting in full session compromise.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Dario Health products are affected:

  • USB-C Blood Glucose Monitoring System Starter Kit Android Applications: Versions 5.8.7.0.36 and prior
  • Dario Application Database and Internet-based Server Infrastructure: All versions

3.2 VULNERABILITY OVERVIEW

3.2.1 EXPOSURE OF PRIVATE PERSONAL INFORMATION TO AN UNAUTHORIZED ACTOR CWE-359

An attacker could expose cross-user Personal Identifiable Information (PII) and personal health information transmitted to the Android device via the Dario Health application database.

CVE-2025-20060 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-20060. A base score of 8.7 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.2.2 IMPROPER OUTPUT NEUTRALIZATION FOR LOGS CWE-117

Unauthenticated log effects metrics gathering incident response efforts and potentially exposes risk of injection attacks (ex log injection).

CVE-2025-23405 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:N).

A CVSS v4 score has also been calculated for CVE-2025-23405. A base score of 6.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:N/SC:N/SI:L/SA:N).

3.2.3 STORAGE OF SENSITIVE DATA IN A MECHANISM WITHOUT ACCESS CONTROL CWE-921

Insecure file retrieval process that facilitates potential for file manipulation to affect product stability and confidentiality, integrity, authenticity, and attestation of stored data.

CVE-2025-24843 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.1 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L).

A CVSS v4 score has also been calculated for CVE-2025-24843. A base score of 5.1 has been calculated; the CVSS vector string is (CVSS4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:L/SC:N/SI:N/SA:N).

3.2.4 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

Lack of encryption in transit for cloud infrastructure facilitating potential for sensitive data manipulation or exposure.

CVE-2025-24849 has been assigned to this vulnerability. A CVSS v3.1 base score of 7.1 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-24849. A base score of 7.5 has been calculated; the CVSS vector string is (CVSS4.0/AV:A/AC:H/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.5 IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79

The Dario Health portal service application is vulnerable to XSS, which could allow an attacker to obtain sensitive information.

CVE-2025-20049 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:L/UI:R/S:C/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-20049. A base score of 7.1 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:H/AT:P/PR:L/UI:A/VC:H/VI:N/VA:N/SC:H/SI:H/SA:H).

Cookie policy is observable via built-in browser tools. In the presence of XSS, this could lead to full session compromise.

CVE-2025-24318 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-24318. A base score of 5.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:H/AT:P/PR:N/UI:A/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N).

3.2.7 EXPOSURE OF SENSITIVE INFORMATION DUE TO INCOMPATIBLE POLICIES CWE-213

The Dario Health Internet-based server infrastructure is vulnerable due to exposure of development environment details, which could lead to unsafe functionality.

CVE-2025-24316 has been assigned to this vulnerability. A CVSS v3.1 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N).

A CVSS v4 score has also been calculated for CVE-2025-24316. A base score of 6.9 has been calculated; the CVSS vector string is (CVSS4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:N/VA:N/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: United States

3.4 RESEARCHER

Noah Cutler and Manuel Del Rio of Accenture reported these vulnerabilities to CISA.

4. MITIGATIONS

Dario Health recommends users update their Dario Health Android mobile application to the latest version. No other actions are required by users.

Dario Health recommends users perform the following mitigations:

  • Update the application from trusted sources.
  • Don't use rooted/jailbroken devices.
  • Avoid public untrusted networks

    For more information contact Dario Health directly.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities, such as:

  • Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls and isolating them from business networks.
  • When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time.

5. UPDATE HISTORY

  • February 27, 2025: Initial Publication

This product is provided subject to this Notification and this Privacy & Use policy.