Contec CMS8000 Contains a Backdoor
This fact sheet details an analysis of three firmware package versions of the Contec CMS8000, a patient monitor used by the U.S. Healthcare and Public Health (HPH) sector. Analysts discovered that an embedded backdoor function with a hard-coded IP address, CWE – 912: Hidden Functionality (CVE-2025-0626), and functionality that enables patient data spillage, CWE – 359: Exposure of Private Personal Information to an Unauthorized Actor (CVE-2025-0683), exists in all versions analyzed.
Please note the Contec CMS8000 may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA’s safety communication, Cybersecurity Vulnerabilities with Certain Patient Monitors from Contec and Epsimed: FDA Safety Communication.
The Contec CMS8000 is used in medical settings in the U.S. and European Union to provide continuous monitoring of a patient’s vital signs. CISA assesses that inclusion of this backdoor in the firmware of the patient monitor can create conditions which may allow remote code execution and device modification with the ability to alter its configuration, introducing risk to patient safety as a malfunctioning patient monitor could lead to an improper response to patient vital signs.